Vinyl siding stain cleaning procedures, chemicals, products:
What methods are recommended by vinyl siding manufacturers and by siding cleaning companies for cleaning dirty, stained, or oxidized vinyl siding?
What goes wrong during siding cleaning procedures: new stains may appear, water can get blown into building walls, shrubs might be killed, ground or nearby waterways might be contaminated if the manufacturer's instructions are not followed.
This article series discusses common causes of stains or discoloration found on vinyl building siding. Distinguishing the type of stain found on siding is useful in deciding on the cause and thus the cure or prevention of staining
InspectAPedia tolerates no conflicts of interest. We have no relationship with advertisers, products, or services discussed at this website.
The Vinyl Siding Institute offers advice on cleaning vinyl siding, recommending basically soft-brush scrubbing, bottom-up, and gives several documents that describe cleaning procedures against which you could review what your cleaning company did:
The Vinyl Siding Institute's Vinyl Siding Installation Manual (includes cleaning advice) https://www.vinylsiding.org/wp-content/uploads/2017/03/2017-Vinyl-Siding-Installation-Manual.pdf Contact:
Jeff Smith, Director, Communications, Vinyl Siding Institute Tel: (202) 587-5109 Email: jsmith@vinylsiding.org Website:
https://www.vinylsiding.org
Excerpt:
Although vinyl siding will get dirty, like anything exposed to the atmosphere, a heavy rain will do wonders to clean it. Or, it’s possible to wash it down with an ordinary garden hose. If neither rain nor hosing does a satisfactory job, follow these simple instructions:
The article also includes suggested cleaners for specific stains. VSI Vinyl Siding Installation Manual (2017).
Watch out: when power-washing siding do not spray "up" such that water is forced underneath or behind the siding or you may cause serious leaks into building walls, in turn cause wet insulation or wall cavity mold contamination.
Also take care that power washing some siding materials such as soft cedar shingles and wood clapboards from too-close a distance can cut into and damage the siding leaving marks that are hard to remove.
Finally, do not power-wash asbestos-cement siding or roofing products aggressively or you may create an asbetstos sludge and later dust hazard.
Watch out: don't use solvents such as acetone, lacquer thinner, or mineral spirits to clean vinyl siding. They can dissolve and permanently-disfigure vinyl and other plastics.
If portions of the siding are discolored even after drying completely there can be various causes, not all of which would be the fault of the cleaner nor cleaning process.
From the example siding cleaning products below you'll see that different manufacturers take varying approaches to the chemistry, use, and suitability of their cleaning products.
Watch out: bleach "stains" on all sorts of siding are a common siding-cleaning issue. Marks may appear as overspray spots that the cleaning contractor didn't notice and that thus left a bleaching cleaner sitting too long on the surface.
I see this problem more often when a siding cleaning contractor, working on area A, allows overspray of their cleaner to splash onto a nearby area B without notice, particularly when working in hot sun or on a dry windy day. The overspray droplets may dry quickly and not even be noticed by the cleaner during the job.
Later as the bleaching cleaner sat longer in some areas than others, yellowing marks of bleached surface may appear.
Marks may also appear as run-down stains, vertical drip marks if during cleaning an upper area of siding the cleaner+cleaned-surface materials ran down to lower siding where it sat too long before rinsing.
Watch out: also for using high-pressure power washers when cleaning siding. Using pressures over 100 psi increase the chances of damage to the siding, power-washer cut marks on softer siding materials (wood), and of blowing water up under the siding and into building walls where it soaks insulation and invites a wall cavity mold contamination problem.
In addition, blowing a power wash up into the siding means that even after the washing contractor thinks they're finished cleaning and rinsing an area of siding there will be additional run-down of cleaning solution that was up inside the siding and that later drains out of siding weep/drain openings where it dwells on the siding surface causing a streaking problem.
This is not an explanation of the spotty stains described earlier in this article: those are probably the result of cleaner overspray that was left on the surface too long.
More examples of stains and problems following cleanign of vinyl siding are
at VINYL SIDING STAIN CLEANING FAQs.
Watch out: when power-washing siding do not spray "up" such that water is forced underneath or behind the siding or you may cause serious leaks into building walls, in turn cause wet insulation or wall cavity mold contamination.
Photo: spotty stains appeared on vinyl siding after it was cleaned by a siding cleaning company.
Just sent you an e mail about my problem and this is the follow up pictures. We will have to reside our entire house if this issue is not solved as it is right by our doorbell and front door and it extensive in this area.
I would really appreciate your ideas as to the caused. I have contacted the company to have it cleaned again and it did not change. - Anonymous by private email 2017/06/10
I can't say for sure what the problem is with the yellowish stains I see on white vertical trim and darker stains I see on grey-beige vinyl siding in your photo. I suspect that a cleaner containing a bleaching agent may have been sprayed or over-sprayed onto a surface while working nearby, then let sit too long without actual washing-off.
I have read but don't know from direct experience that some vinyl siding, particularly darker colors, vary in chemical composition and thus in their susceptability to chalking and staining before and as a result of washing.
I also read that some less-costly vinyl siding products are a laminated product whose surface color is just a thin layer that is more-easily damaged by cleaning agents.
That is to say some siding cleaners blame the product for some stain issues and suggest that you check the siding warranty.
I would try some local spot cleaning in a less-obtrusive location to see if a household cleaner, scrubbing, or a cleaner suggested by your contractor get rid of the trouble. If the siding surface was oxidized and chalky you may be able to reduce the yellow blotches by gentle scrubbing with a properly-diluted bleaching cleaner.
General advice when cleaning a surface to try to remedy a dirt or stain problem: always try the least-aggressive method and least-aggressive cleaner (start with plain water or plain water and a power washer or scrub brush) first.
When stains appear after washing, sometimes waiting a few days for the surface to dry thoroughly may show that the stains mostly disappear; (On the other hand I've had complaints from readers that some stucco siding products had stains that reappeared every time it rained.)
When the coating on siding is oxidized some siding cleaners use a cleaning solution that contains potassium hydroxide (KOH) or alternatively a butyl-based cleaner to emulsify the oxidized particles on the siding surface.
Basically you'd need at least a cleaner that includes a "surface debris surfactant" to clean such siding.
Watch out: I am NOT prescribing ANY specific treatment for your home, as no one in their right mind could try to do so based on one photo and an e-text message. But these are options among those you might want to discuss with your siding cleaning contractor.
So IF the stains were - by luck - principally in the oxidized layer of a siding surface, then it may be possible to remove or significantly-reduce them by this more labor-intensive (and more expensive) cleaning process.
Oxidized siding surfaces is a condition you can usually detect by noticing a chalky deposit on the surface that can be smeared with a thumb or scrubbed away with a soft brush and water or water + detergent.
Oxidation of the siding surface (or of paints on surfaces) is a result of sunlight (UV exposure) that breaks down chemicals in the surface.
An experienced siding cleaning company might know to warn the customer if they see oxidized siding surfaces that simple "cleaning" won't be effective and that a more costly and troublesome "siding restoration" procedure involving more-careful cleaning will be needed.
If I were the contractor I'd warn, in writing that oxidized siding presents special and extra risks of blotching, spotting, streaking stains following cleaning efforts.
I also warn that on some older aluminum siding that was oxidized I've seen cleaning remove all of the paint pigment leaving shiny aluminum!
Bottom line: my OPINION is that bleach-caused spots on vinyl siding and trim won't be removed once a bleaching agent sat long enough to cause yellowing and that the only option to fix what is primarily a cosmetic problem is either painting the siding or re-siding the side of the home with the most significant damage.
Here we describe cleaners, their main ingredients, and the basic chemical properties of those ingredients. Some popular cleaning approaches and products used for vinyl siding cleaning include the following:
Following the use of a siding cleaner with low pressure washing or soft brushing is usually effective.
Never use these products on vinyl siding:
Watch out: some of these are dangerous chemicals and must be used following the manufacturer's directions and with proper protective gear.
...
Continue reading at VINYL SIDING STAINS - topic home, or select a topic from the closely-related articles below, or see the complete ARTICLE INDEX.
Or see VINYL SIDING STAIN CLEANING FAQs - questions & answers about stains appearing after cleaning vinyl siding
Or see these
VINYL SIDING CLEANING METHODS at InspectApedia.com - online encyclopedia of building & environmental inspection, testing, diagnosis, repair, & problem prevention advice.
Or see this
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Questions & answers or comments about Vinyl Siding on Buildings, Installation, Inspection Methods, Defects, Diagnosis & Repair.
Try the search box just below, or if you prefer, post a question or comment in the Comments box below and we will respond promptly.
Search the InspectApedia website
Note: appearance of your Comment below may be delayed: if your comment contains an image, photograph, web link, or text that looks to the software as if it might be a web link, your posting will appear after it has been approved by a moderator. Apologies for the delay.
Only one image can be added per comment but you can post as many comments, and therefore images, as you like.
You will not receive a notification when a response to your question has been posted.
Please bookmark this page to make it easy for you to check back for our response.
Our Comment Box is provided by Countable Web Productions countable.ca
In addition to any citations in the article above, a full list is available on request.